Portable Electric IV Infusion Pump

In stock

Place of Origin: China
Brand Name: Healicom
Model Number: HSA513
Power Source: Electric
Warranty: 1 Year
Material: Metal, Plastic
Quality Certification: ce
Instrument classification: Class II
Infusion precision: ±5%
Color: Black and White
Infusion volume error: ±5%
Bolus rate(mL/h)and error: 200~1200 continuously adjustable;±20%;
Occlusion alarm level(pressure): 50kPa±25kPa
Max. pressure from the infusion set: More than 120kPa
Pause overtime alarm time: 1 min 50 s ~ 2 min
Recoverable alarm: 1 min 50 s ~ 2 min
service life: 5 year

Description

Portable Electric IV Infusion Pump

H41a606a7254244a19c0b62d5b79b3bfbe
Technical Specifications
 
Infusion Rate range(mL/h)
1-1200;
1mL/h~99.9mL/h range adjustable in 0.1mL/h steps or 1mL/h steps;
100.0mL/h~1200mL/h adjustable in 1mL/h steps only;
Infusion precision
±5% (correctly calibrated and operated in a stable environment according to the requirement of 10)
Pre-set of infusion volume(mL)
0~9999, 1mL steps
Infusion volume error
±5%; (correctly calibrated and operated in a stable environment according to the requirement of 10)
Bolus rate(mL/h) and error
200~1200 continuously adjustable; ±20%;
Occlusion alarm level (pressure)
50kPa±25kPa
Max. pressure from the infusion set
More than 120kPa
 
 
 

Triggering time of occlusion alarm and bolus dose

Minimum flow rate: When the occlusion alarm pressure is 50kPa±25kPa, Less than 13 minutes.
Intermediate flow rate: When the occlusion alarm pressure is 50kPa±25kPa, Less than 14 seconds.
Occlusion dose does not beyond 0.3mL
(use Jerry infusion set, chokepoint is made at the tail end of the infusion set.)
Keep vein open(KVO)rate
KVO rate of 3mL/h with flow rate higher than 10mL/h, 1mL/h with flow rate lower or equals to 10mL/h and higher than 1 mL/h, 0.1 mL/h with flow rate lower than 1 mL/h, accuracy error+ 20%
Recoverable alarm
1 min 50 s ~ 2 min
Pause overtime alarm time
1 min 50 s ~ 2 min
Classification
Type CF, Class II continuous volumetric infusion pump with internal battery; IPX4 (sealed drip-proof device)
He9e48778ab8149399525e993ca5869413
H2224d36ad8a74b0ca5d4da307bc76258Q
Visit our Facebook page: Tista Bangladesh

প্রশ্ন-উত্তর পর্ব

যে কোন তথ্য জানতে এখানে প্রশ্ন করুন


 

Main Menu